1.1 Operating Principle
Oxidative destruction of microorganisms
Thermal decomposition of cellular components
Typical temperature range: 160°C-190°C (320°F-374°F)
1.2 Time-Temperature Relationships
Where: t = exposure time (minutes) D₁₆₀ = 60 minutes (Bacillus atrophaeus) z-value = 20°C
Validation Standards
ISO 20857:2013 compliance requirements
Heat distribution testing (±5°C uniformity)
2.1 Temperature Ranges
Process | Typical Range | Peak Efficiency |
---|---|---|
Sausais siltums | 160-190°C | 170°C/60min |
Steam Autoclave | 121-135°C | 132°C/4min |
2.2 Penetration Capacity
Dry heat: 0.5°C/cm depth penetration rate
Steam: 2.3°C/cm depth penetration rate
2.3 Material Compatibility
Dry heat preferred for:
Anhydrous powders
Glassware
Metal instruments with crevices
Steam required for:
Rubber components
Wrapped instrument kits
Culture media
3.1 Dry Heat Oven Setup
Pre-heat to 140°C for 30 minutes
Ramp to target temperature (≤5°C/min)
Maintain thermal equilibrium (±3°C)
Cooling phase protocols (natural vs forced)
3.2 Cycle Timing Guidelines
Temperature | Minimum Time | Typical Applications |
---|---|---|
160°C | 120 minutes | Glass pipettes |
170°C | 60 minutes | Surgical drills |
180°C | 30 minutes | Orthopedic implants |
190°C | 6 minutes | Emergency instruments |
4.1 Biological Indicators
Bacillus atrophaeus spore strips (10⁶ CFU)
Placement: Coldest oven location
Incubation: 56°C for 48 hours
4.2 Physical Monitoring
Thermocouple mapping (9-point test)
Heat penetration studies
Airflow velocity verification (>2 m/s)
Dry heat sterilization remains essential for moisture-sensitive instruments, requiring higher temperatures but offering superior material compatibility for specific applications. Successful implementation demands rigorous validation of both time-temperature parameters and heat distribution characteristics. Facilities must choose sterilization methods based on load composition rather than convenience.
Q: How long does sterilization with a dry heat oven typically take? A: Standard cycles range from 6 minutes at 190°C to 2 hours at 160°C. Actual duration depends on load mass and oven validation data.
Q: Can dry heat achieve the same sterility assurance level as autoclaves? A: Both methods achieve 10⁻⁶ SAL when properly validated. Dry heat requires longer cycles due to slower microbial destruction kinetics.
Q: Why are dry heat temperatures higher than autoclave settings? A: Moisture absence necessitates higher thermal energy input. Steam’s latent heat transfer enables lower temperatures (121°C vs 160°C).
Q: How to verify dry heat sterilization effectiveness? A: Combine biological indicators with chemical integrators and thermometric data loggers. Quarterly validation using loaded chamber studies is mandatory.
Sterilizācija ir būtisks infekciju kontroles prakses elements gan veselības aprūpes iestādēs, gan laboratorijās. Autoklāvā sterilizācija ar tvaiku ir efektīva daudzos gadījumos, bet nav optimāla, ja
Medicīnas profesionāļi visā pasaulē uzticas autoklāviem kā būtiskiem instrumentiem, lai nodrošinātu medicīnas instrumentu drošību un sterilitāti. Medicīnas ierīču izplatītājiem un iepirkumu speciālistiem ir jāizprot autoklāvu darbības principi un priekšrocības.
Veselības aprūpes sasniegumi nosaka to, ka ir svarīgi, lai medicīniskie instrumenti vienmēr būtu droši un sterili. Medicīnas ierīču izplatītājiem, tirgotājiem un iepirkumu speciālistiem ir jāizprot sterilizācijas metodes, lai varētu efektīvi darboties. Portāls
Medicīnisko ierīču un instrumentu sterilizācijai ir nepieciešami autoklāvi, jo tie aizsargā pacientu drošību un veicina atbilstību normatīvajiem standartiem. Vismodernākās autoklāvu sistēmas joprojām saskaras ar zināmiem ierobežojumiem. Medicīnas ierīču izplatītāji,
Medicīnisko ierīču ražošanā autoklāvs ir svarīgs aprīkojums, jo tas nodrošina konsekventu sterilizāciju dažādiem instrumentiem un materiāliem. Autoklāvos tiek izmantots augsta spiediena piesātināts tvaiks, lai iznīcinātu baktērijas, vīrusus, sēnītes un citus patogēnus.
Autoklāvs kalpo kā ļoti svarīga ierīce veselības aprūpē un zinātnē, konsekventi sterilizējot instrumentus un materiālus. Medicīniskā aprīkojuma izplatītājiem, kā arī tirgotājiem un iepirkumu speciālistiem ir jānodrošina autoklāvu darbība.