Avtoklavni stroji: Kako delujejo, vrste in pravilne tehnike sterilizacije

Avtoklavi služijo kot ključna orodja za sterilizacijo v laboratorijih ter zdravstvenih in industrijskih okoljih. Ti sistemi uporabljajo paro pod pritiskom pri temperaturah, višjih od 121 °C (250 °F), da odstranijo bakterije, viruse in spore z učinkovitostjo 99,9999%. Vodnik zajema bistvena načela delovanja avtoklavov, ključne konstrukcijske značilnosti in protokole, ki zagotavljajo stabilne rezultate sterilizacije ter optimalno življenjsko dobo opreme.

1. How Autoclave Machines Work

The process of sterilizing materials in autoclaves depends on the interaction between pressure and temperature.

  1. Phase 1 – Air Removal

  • In gravity displacement (Class N) sterilization, steam drives air out through the drain vent.

  • The vacuum pump in pre-vacuum Class B autoclaves eliminates 99.9% of air before steam injection begins.

  1. Phase 2 – Sterilization

  • During the standard sterilization cycle, the autoclave maintains 121°C and 15 psi pressure for a period between 15 and 30 minutes.

  • Fast sterilization cycles run at 134°C under 30 psi pressure for durations between 3 and 10 minutes.

  1. Phase 3 – Drying

  • Post-vacuum phase removes residual moisture

  • Typical drying time: 20-45 minutes at 80-90°C

Key Formula: The sterilization effectiveness (D-value) is calculated using: [ D{121} = \frac{\text{Exposure Time}}{\log{10}(N_0/N)} ] (N_0) stands for the initial number of microbes and (N) represents the number of microbes that survived.

2. Autoclave Components & Specifications

KomponentaFunkcijaCritical Specifications
KomoraHolds items for sterilization316L stainless steel, ≥1.5mm thick
Generator pareProduces pure steamOutput: 3-6 kg/h per 100L chamber
Nadzorni sistemManages cycle parameters±0.5°C accuracy, 10+ preset programs
Varnostni ventilPrevents over-pressurizationASME Section VIII compliant

Združljivost materialov:

  • Autoclavable: Stainless steel, borosilicate glass, polypropylene

  • Non-autoclavable: PVC, polystyrene, low-density polyethylene

3. Autoclave Validation & Testing

  1. Fizično spremljanje

  • Chart recorders: Verify time-temperature-pressure curves

  • Data loggers: Track 3D chamber heat distribution

  1. Kemični kazalniki

  • Class 1: Process indicators (tape color change)

  • Class 4: Multi-variable integrators

  1. Biološki kazalniki

  • Use Geobacillus stearothermophilus spores (10⁶ population)

  • Incubation: 56-60°C for 24-48 hours

Acceptance Criteria:

  • No growth in test vials

  • Temperature uniformity ≤±1°C across chamber

4. Step-by-Step Operating Procedure

  1. Nalaganje:

  • Arrange items with 2.5cm spacing between objects

  • Position pouches printed-side up

  1. Izbira cikla:

  • Liquids: Slow exhaust gravity cycle

  • Wrapped instruments: Pre-vacuum cycle

  1. Razkladanje:

  • Wait until internal pressure reaches 0 psi

  • Use thermal gloves for hot items

  1. Dokumentacija:

  • Record cycle number, parameters, operator ID

5. Maintenance Schedule

TaskFrekvencaTools/Agents
Door gasket inspectionDailyVisual check, leak detector
Chamber cleaningWeeklyNon-abrasive citric acid solution
Safety valve testMesečnoCalibrated pressure gauge
KalibracijaLetnoNIST-traceable thermometer

Descaling Protocol:

  1. Fill chamber with 5% acetic acid solution

  2. Run 30-minute cycle at 100°C

  3. Rinse 3× with deionized water


Zaključek

Autoclave machines that undergo proper operation and maintenance deliver dependable sterilization which is indispensable for controlling infections and ensuring research results remain valid. Consistent steam sterilization results can be obtained when users follow validated cycles along with regular biological monitoring and manufacturer instructions. Technological advancements such as cloud monitoring systems and automated validation tools will improve autoclave performance tracking in future years.


Pogosta vprašanja

Q1: Can autoclaves sterilize liquids? Specialized liquid cycles with slow exhaust protect against boiling over during autoclave sterilization. Always fill containers ≤75% capacity.

Q2: How often should autoclave validation be performed? The ISO 17665 standard mandates quarterly validation of medical devices. Laboratories need to perform autoclave validation every year or following significant repairs.

Q3: Why does my autoclave take so long to reach temperature? Common causes include insufficient steam generation (check boiler capacity) or air leaks (test door gasket integrity).

Q4: What water quality is required for autoclaves? Prevent mineral buildup in autoclaves by using ASTM Type II water with conductivity levels below 5 µS/cm. Distilled or RO water is acceptable.

Q5: How long do autoclave cycles typically last? Autoclave standard cycles last between 20 to 90 minutes based on load type. Unwrapped items achieve sterilization within 3 minutes through fast flash cycles.

Delite:

Več prispevkov

Pošljite nam sporočilo

Če želite izpolniti ta obrazec, v brskalniku omogočite JavaScript.
Keling-Medical-Autoclave-030

avtoklav acqua

Postopek avtoklaviranja je bistvena praksa sterilizacije, ki se uporablja v medicinskih, laboratorijskih in raziskovalnih ustanovah za zaščito steklene posode in instrumentov z učinkovito sterilizacijo. Visokotlačna para pri tem postopku odstrani patogene.

Preberite več "
Keling-Medical-Autoclave-032

avtoklav

Postopek avtoklaviranja je bistvena praksa sterilizacije, ki se uporablja v medicinskih, laboratorijskih in raziskovalnih ustanovah za zaščito steklene posode in instrumentov z učinkovito sterilizacijo. Visokotlačna para pri tem postopku odstrani patogene.

Preberite več "

zobozdravstveni avtoklav

Postopek avtoklaviranja je bistvena praksa sterilizacije, ki se uporablja v medicinskih, laboratorijskih in raziskovalnih ustanovah za zaščito steklene posode in instrumentov z učinkovito sterilizacijo. Visokotlačna para pri tem postopku odstrani patogene.

Preberite več "

avtoklavni laboratorij

Postopek avtoklaviranja je bistvena praksa sterilizacije, ki se uporablja v medicinskih, laboratorijskih in raziskovalnih ustanovah za zaščito steklene posode in instrumentov z učinkovito sterilizacijo. Visokotlačna para pri tem postopku odstrani patogene.

Preberite več "

avtoklava ceye

Postopek avtoklaviranja je bistvena praksa sterilizacije, ki se uporablja v medicinskih, laboratorijskih in raziskovalnih ustanovah za zaščito steklene posode in instrumentov z učinkovito sterilizacijo. Visokotlačna para pri tem postopku odstrani patogene.

Preberite več "

deli avtoklava

Postopek avtoklaviranja je bistvena praksa sterilizacije, ki se uporablja v medicinskih, laboratorijskih in raziskovalnih ustanovah za zaščito steklene posode in instrumentov z učinkovito sterilizacijo. Visokotlačna para pri tem postopku odstrani patogene.

Preberite več "