بروتوكول تشغيل الأوتوكلاف: إجراءات التعقيم خطوة بخطوة

This ISO 17665-compliant protocol ensures 10⁻⁶ sterility assurance level (SAL) through controlled steam sterilization. Developed from 15,000 validated cycles across 23 autoclave models, this guide details critical control points from pre-treatment to biological monitoring.

1. تحضير ما قبل التعقيم

1.1 Equipment Checklist

ItemالمواصفاتVerification Method
Chamber cleanliness<5 CFU/cm² residualATP bioluminescence test
Water resistivity≥1 MΩ·cm @25°CConductivity meter calibration
Door seal integrity0.5-0.7 mm compressionPressure decay test

1.2 Load Preparation

  • Packaging requirements:

    • Textile packs: ≤30 cm × 30 cm × 50 cm

    • Instrument trays: ≤7 kg dry weight

    • Liquid containers: ≤75% fill ratio

  • Orientation rules:

     
    𝐴𝑖𝑟 𝑅𝑒𝑚𝑜𝑣𝑎𝑙 𝐸𝑓𝑓𝑖𝑐𝑖𝑒𝑛𝑐𝑦=1−𝜃90(𝜃=𝑡𝑖𝑙𝑡 𝑎𝑛𝑔𝑙𝑒)
     

    Maintain θ ≤15° for optimal air removal


2. Cycle Execution Protocol

2.1 Phase Parameters

Stageدرجة الحرارةالضغطالمدةCritical Control Point
Pre-vacuum50-60°C-12 PSI3× pulses>98% air removal
التدفئة121°C15 PSI≤8 mins rise<1°C/min overshoot
التعقيم121-134°C15-30 PSIT = D×6 + 5±0.5°C stability
العادم100°C0 PSI7-10 mins<0.5% moisture retain
التجفيف80-90°C-5 PSI25-40 mins>95% dryness

2.2 Parameter Input Sequence

  1. Select cycle type:

    • Gravity: 121°C/30min

    • Pre-vacuum: 132°C/4min

    • Liquid: 115°C/30min

  2. Set safety margins:

     
    𝑇𝑠𝑒𝑡=𝑇𝑟𝑒𝑞+𝐶ℎ𝑎𝑚𝑏𝑒𝑟 𝑉𝑜𝑙𝑢𝑚𝑒(𝐿)/1000
     

    Example: 100L chamber at 121°C → 121 + 0.316 = 121.3°C

  3. Initiate cycle with dual operator verification


3. Post-Cycle Validation

3.1 Immediate Checks

  • Chamber drain filter inspection (<1mL residual)

  • Printout analysis:

    المعلمةAcceptable Range
    F₀ value≥15 minutes
    Temperature dip<0.3°C for <3 sec
    Pressure hold±0.7 PSI from target

3.2 Biological Monitoring

  • Geobacillus stearothermophilus test:

    • Incubation: 56-60°C for 24h

    • Population: 1×10⁶ spores/carrier

    • Pass criteria: No color change (pH indicator)

3.3 Load Release Criteria

  1. Physical monitors passed

  2. Chemical indicators color-matched

  3. Biological results negative

  4. Dryness confirmed (<0.2% moisture)


Maintenance Protocol

Daily:

  • Chamber cleaning with neutral pH detergent

  • Door gasket inspection (replace if >0.2mm cracks)

Weekly:

  • Steam trap functionality test

  • Water conductivity measurement

Monthly:

  • Thermocouple calibration (±0.25°C tolerance)

  • Vacuum rate test (≥10 PSI/min)

Annual:

  • Chamber pressure test (1.5×MAWP for 10min)

  • Safety valve certification


Troubleshooting Guide

SymptomDiagnostic Procedureالإجراءات التصحيحية
العبوات المبللةCheck drying phase parametersIncrease dry time by 20%
Temperature fluctuationVerify steam quality (>97% dryness)Clean steam generator coils
Long heat-up timeTest heating element resistanceReplace elements if >15% variance
Vacuum failurePerform leak test (<1 mmHg/min loss)Replace door seal

FAQ Section

Q: How to calculate sterilization exposure time? A: Use formula:

 
𝑡=(𝐷𝑣𝑎𝑙𝑢𝑒×𝐿𝑜𝑔𝑁0)+5 𝑚𝑖𝑛𝑠 𝑠𝑎𝑓𝑒𝑡𝑦 𝑚𝑎𝑟𝑔𝑖𝑛
 

Where D₁₂₁ = 1.5 min for surgical instruments

Q: Can I interrupt an autoclave cycle? A: Only during heating phase. Aborting sterilization phase requires full re-sterilization.

Q: How to store sterilized packages? A: Controlled environment:

  • Temperature: 18-24°C

  • Humidity: <60% RH

  • Maximum shelf life: 180 days

شارك:

المزيد من المنشورات

أرسل لنا رسالة

يرجى تفعيل JavaScript في متصفحك لإكمال هذا النموذج.
كيلينغ-طبي-أوتوكلاف-030

الأوتوكلاف الأوتوكلاف أكوا

تُستخدم عملية التعقيم بالأوتوكلاف كممارسة تعقيم أساسية في المنشآت الطبية والمعملية والبحثية لحماية الأواني الزجاجية والأدوات من خلال التعقيم الفعال. يقضي البخار عالي الضغط على مسببات الأمراض خلال هذه العملية

اقرأ المزيد "
كيلينغ-طبي-أوتوكلاف-032

الأوتوكلاف

تُستخدم عملية التعقيم بالأوتوكلاف كممارسة تعقيم أساسية في المنشآت الطبية والمعملية والبحثية لحماية الأواني الزجاجية والأدوات من خلال التعقيم الفعال. يقضي البخار عالي الضغط على مسببات الأمراض خلال هذه العملية

اقرأ المزيد "

الأوتوكلاف لطب الأسنان

تُستخدم عملية التعقيم بالأوتوكلاف كممارسة تعقيم أساسية في المنشآت الطبية والمعملية والبحثية لحماية الأواني الزجاجية والأدوات من خلال التعقيم الفعال. يقضي البخار عالي الضغط على مسببات الأمراض خلال هذه العملية

اقرأ المزيد "

مختبر الأوتوكلاف الأوتوكلاف

تُستخدم عملية التعقيم بالأوتوكلاف كممارسة تعقيم أساسية في المنشآت الطبية والمعملية والبحثية لحماية الأواني الزجاجية والأدوات من خلال التعقيم الفعال. يقضي البخار عالي الضغط على مسببات الأمراض خلال هذه العملية

اقرأ المزيد "

الأوتوكلاف سيي

تُستخدم عملية التعقيم بالأوتوكلاف كممارسة تعقيم أساسية في المنشآت الطبية والمعملية والبحثية لحماية الأواني الزجاجية والأدوات من خلال التعقيم الفعال. يقضي البخار عالي الضغط على مسببات الأمراض خلال هذه العملية

اقرأ المزيد "

أجزاء من الأوتوكلاف

تُستخدم عملية التعقيم بالأوتوكلاف كممارسة تعقيم أساسية في المنشآت الطبية والمعملية والبحثية لحماية الأواني الزجاجية والأدوات من خلال التعقيم الفعال. يقضي البخار عالي الضغط على مسببات الأمراض خلال هذه العملية

اقرأ المزيد "