[Insert Schematic Diagram Here: Simplified autoclave cross-section with numbered parts]
The autoclave sterilization chamber is made from SUS 316L stainless steel which ranges from 3mm to 6mm thickness and meets ASTM A240 standards.
The sterilization chamber design options include circular or rectangular models that function at pressures up to 40 psi.
용량: 25L (dental) to 1,500L (hospital CSSD)
Sealing Mechanism:
Silicone gasket (Shore A 50–60 hardness)
Pneumatic or manual locking (3–5 locking bolts)
안전 기능:
Pressure interlock prevents opening above 0 psi
Thermal shield reduces surface temp to <50°C
구성 요소 | Specification | 기능 |
---|---|---|
Jacketed Chamber | 10mm insulation layer | Prevents heat loss |
증기 발생기 | 4–15 kW output | Produces 97% dry saturated steam |
Pressure Sensor | 0–50 psi range ±0.1% accuracy | Monitors cycle progress |
Thermostatic Trap | Opens at 100°C ±2°C | Removes condensate |
Vacuum Pump: 60–100 mbar residual pressure
The vacuum system eliminates 99.9% of air within two minutes following EN 13060 standards.
애플리케이션: Textiles, wrapped instruments, porous loads
UI Types:
Basic: LED indicators + mechanical timers
Advanced: 7″ touchscreen with data export
Validation-Ready Features:
USB/Bluetooth cycle data download
Password-protected parameter adjustment
Critical Component Lifespan:
Part | Replacement Interval | Failure Consequences |
---|---|---|
도어 개스킷 | 800–1,200 cycles | Steam leaks, aborted cycles |
Steam Generator Coil | 5–8 years | Reduced steam quality |
Solenoid Valves | 3–5 years | Pressure regulation failures |
Material Selection Guide:
애플리케이션 | Preferred Material | Rationale |
---|---|---|
Chamber Interior | SUS 316L | Chloride corrosion resistance |
Steam Lines | ASTM A106 Gr.B | High-temperature durability |
Electrical Contacts | Silver-Nickel Alloy | Oxidation prevention |
When purchasing autoclaves distributors need to validate these parameters.
Chamber Certification:
The autoclave chamber needs to have an ASME BPVC Section VIII stamp for pressure vessels certification.
Passivation certificate (ASTM A967)
Steam Quality Metrics:
Dryness fraction: ≥97%
Non-condensable gases: ≤3.5% (EN 285 standard)
Vacuum Performance:
The air removal rate must result in ≤1% residual air to pass the Bowie-Dick test.
에너지 효율성:
The steam usage rate stands at no more than 5 kilograms for every 100 liters of chamber volume.
예방 일정:
Task | 빈도 | Tools Required |
---|---|---|
Gasket Inspection | Weekly | Leak detection spray |
Calibration Check | 분기별 | NIST-traceable pressure gauge |
Chamber Descaling | 500 cycles | 10% citric acid solution |
일반적인 문제 해결:
Problem: Incomplete drying 솔루션: Clean strainer in exhaust line
Problem: Error code “E3” (low water) 솔루션: Check float switch alignment
Medical buyers prioritize autoclaves meeting:
ISO 13485: Quality management for medical devices
EN 13060: Small steam sterilizer standard
21 CFR 880.6860: FDA performance criteria
Documentation Checklist:
Material test certificates (MTC) for chamber
Pressure equipment directive (PED) compliance
Electromagnetic compatibility (EMC) test reports
IoT 통합:
Real-time part wear monitoring via strain gauges
Sustainable Materials:
Graphene-coated chambers for enhanced corrosion resistance
Additive Manufacturing:
3D-printed steam nozzles optimized for flow dynamics
Medical equipment distributors who understand autoclave components gain a competitive edge. Buyers achieve global standard sterilization solutions with minimal downtime by thoroughly evaluating chamber materials and steam systems alongside control interfaces. Predictive maintenance strategies in healthcare facilities will drive the market for smart diagnostic modular autoclaves as the dominant force in the $3.2 billion sterilization equipment sector.
Q1: How often should door seals be replaced? A: Every 1,200 cycles or if visible cracks/compression set exceeds 15%.
Q2: Can chamber material impact sterilization efficacy? A: Yes. Pitted surfaces trap contaminants – SUS 316L maintains smoothness for 10,000+ cycles.
Q3: What’s the cost difference between basic and advanced control panels? A: Touchscreen models cost 25–40% more but reduce validation time by 70%.
Q4: How to verify vacuum pump performance? A: Conduct daily air leakage tests – pressure rise ≤1 mbar/min indicates proper function.
Q5: Are stainless steel trays mandatory? A: Aluminum trays are acceptable but require passivation every 6 months (ASTM A967).
고객에게 다음을 제공하세요. technologically advanced autoclaves built for precision and compliance.
📧 이메일: inquiry@shkeling.com 📱 WhatsApp: +86 182 2182 2482 🌐 Explore Models: Industrial & Portable Autoclaves
왜 우리를 선택해야 할까요?
Certified Excellence: CE, ISO 13485, and FDA-compliant designs
Custom Engineering:
Dual-door pass-through configurations
Left/right-handed door options
B2B Benefits:
Bulk pricing for 10+ units
Spare parts inventory access
Multilingual installation guides
기간 한정 혜택: ✅ Free component lifetime calculator with quote request ✅ Priority technical support for distributors