
Key Microbial Destruction Metrics:
D-value represents the time duration necessary to decrease microbial population by 90% when exposed to certain temperatures
Example: Geobacillus stearothermophilus D₁₂₁ = 1.5-2.0 minutes
Z-value represents the temperature shift required to achieve a 90% change in the D-value.
Typical z = 10°C for steam sterilization
Sterility Assurance Level (SAL): [ SAL represents a probability of one viable microorganism surviving among one million attempts.]
Parameter | Requirement | Tolerance | Measurement Tool |
---|---|---|---|
Temperatūra | 121°C (Gravity) / 134°C (Pre-Vacuum) | ±1°C | NIST-calibrated thermocouples |
Slėgis | 15 psi (121°C) / 30 psi (134°C) | ±5% | Bourdon tube gauge |
Ekspozicijos laikas | 15-30 minutes (121°C) | +0/-1 minute | Digital timer |
Steam Quality | 97% dry saturated steam | Non-condensable gases <3.5% | Karl Fischer titration |
Material Compatibility Guidelines:
Textile porous loads need long drying durations exceeding 45 minutes when heated to 80°C.
For liquids, implement a slow exhaust cycle to stop boiling over.
Pre-vacuum cycles with more than three pulse phases are necessary for hollow devices.
Phase 1: Preparation
Cleaning
Remove bioburden with enzymatic detergents (pH 6.5-7.5)
Perform a rinse with ASTM Type I water which maintains conductivity below 1 μS/cm.
Pakuotė
Use medical-grade paper/plastic pouches (ISO 11607 compliant)
Apply 5mm seal margins to packages and attach corresponding LOT identifiers.
Pakrovimas
Maintain 2.5cm spacing between items
Position pouches print-side up
Phase 2: Sterilization Cycle
Oro šalinimas
Gravity displacement: 3-5 air purges
Pre-vacuum: 0.8-1.0 bar negative pressure
Šildymo etapas
Ramp rate: ≤1°C/second to prevent thermal shock
Hold until coldest chamber point reaches setpoint
Ekspozicijos laikas
Start timing when all probes ≥setpoint temperature
Minimum dwell time based on load type:
Metal instruments: 3 mins @134°C
Textiles: 15 mins @121°C
Cooling/Drying
Gradual pressure release (0.5 bar/min)
Vacuum-assisted drying to ≤3% moisture content
Phase 3: Post-Process Handling
Aušinimas
Ambient cooling racks (stainless steel 304)
Time-out period: 30 mins before handling
Saugykla
Controlled environment: 23°C ±2°C, 30-60% RH
Products remain stable for 180 days once wrapped but last only 24 hours when unwrapped.
Physical Monitoring:
Data loggers utilize 12 or more thermocouples for mapping cold spots within the chamber.
Pressure transducers measure 0-5 bar pressure levels and deliver results with 0.1% accuracy.
Cheminiai rodikliai:
Class | Purpose | Pass Criteria |
---|---|---|
1 | Process exposure | Color change from tan→black |
4 | Multi-variable integration | Reference mark attainment |
5 | Cycle-specific verification | All parameters met |
Biologiniai rodikliai:
Spore strips: Geobacillus stearothermophilus ATCC 7953
Incubation: 56-60°C for 7 days (extended readout)
The acceptable standard requires no positive growth after 20 tests.
Revalidation Requirements:
Annual requalification (ISO 17665)
After major repairs or relocation
When changing load configurations
Failure Mode | Root Cause | Corrective Action |
---|---|---|
Wet packs | Insufficient drying time | Extend dry phase by 50% |
Cold spots | Air entrapment | Add porous load adapters |
Cycle aborts | Steam quality issues | Install duplex water softeners |
Biological positives | Insufficient exposure time | Increase dwell time by D-value×6 |
The autoclave sterilization process requires precise adherence to validated time-temperature-pressure parameters through physical, chemical, and biological monitoring. Facilities achieve dependable microbial inactivation and maintain material integrity by operating ISO 17665-compliant cycles and following strict equipment calibration schedules. Real-time wireless monitoring and AI-powered cycle optimization represent emerging technologies that will transform sterilization best practices during the next decade.
Q1: What duration is required for completing a standard autoclave sterilization process? A: Standard autoclave cycles take between 40 to 90 minutes and incorporate heating along with exposure and drying stages. Emergency instruments require flash cycles that complete within 3 to 10 minutes.
Q2: What temperature kills spores in autoclaves? A: Geobacillus stearothermophilus spores need a minimum of 121°C for 15 minutes to be destroyed. At 134°C the necessary exposure time is reduced to just 3 minutes.
Q3: Can you autoclave plastic items? A: The only plastic materials suitable for autoclave processes are heat-resistant types such as polypropylene (PP) and PTFE. Check material SDS for maximum tolerated temperatures.
Q4: How often should biological testing be performed? A: Medical devices must undergo weekly testing according to AAMI ST79 requirements. Testing intervals in labs should be set to either monthly or after every 40 cycles.
Q5: Why is steam quality critical in autoclaving? A: Steam below 97% dryness level decreases heat transfer effectiveness and results in wet packs. Monitor steam quality quarterly via thermochemical tests.
Autoklavavimo procesas yra esminė sterilizacijos praktika, taikoma medicinos, laboratorijų ir mokslinių tyrimų įstaigose, siekiant veiksmingai sterilizuoti stiklo gaminius ir instrumentus. Aukšto slėgio garai pašalina patogenus per šį procesą.
Autoklavavimo procesas yra esminė sterilizacijos praktika, taikoma medicinos, laboratorijų ir mokslinių tyrimų įstaigose, siekiant veiksmingai sterilizuoti stiklo gaminius ir instrumentus. Aukšto slėgio garai pašalina patogenus per šį procesą.
Autoklavavimo procesas yra esminė sterilizacijos praktika, taikoma medicinos, laboratorijų ir mokslinių tyrimų įstaigose, siekiant veiksmingai sterilizuoti stiklo gaminius ir instrumentus. Aukšto slėgio garai pašalina patogenus per šį procesą.
Autoklavavimo procesas yra esminė sterilizacijos praktika, taikoma medicinos, laboratorijų ir mokslinių tyrimų įstaigose, siekiant veiksmingai sterilizuoti stiklo gaminius ir instrumentus. Aukšto slėgio garai pašalina patogenus per šį procesą.
Autoklavavimo procesas yra esminė sterilizacijos praktika, taikoma medicinos, laboratorijų ir mokslinių tyrimų įstaigose, siekiant veiksmingai sterilizuoti stiklo gaminius ir instrumentus. Aukšto slėgio garai pašalina patogenus per šį procesą.
Autoklavavimo procesas yra esminė sterilizacijos praktika, taikoma medicinos, laboratorijų ir mokslinių tyrimų įstaigose, siekiant veiksmingai sterilizuoti stiklo gaminius ir instrumentus. Aukšto slėgio garai pašalina patogenus per šį procesą.